Model Number 4.75MM BC KNOTLESS SWIVELOCK
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for evaluation; no pictures were provided.The most likely cause for the reported failure is user error.Per dfu-0087-11.G.Precautions.5.Pushlock and swivelock suture anchor only: insert the driver into the bone socket until the anchor body makes contact with the bone.Preview and adjust suture tension, if necessary.Tension will not increase during the final advancement of the anchor body.
 
Event or Problem Description
It was reported that a biocomposite knotless swivelock, ar-2324kbcc, was being used in a procedure.When putting the anchor into the punched bone tunnel, the two fibertapes that were in the eyelet, ripped through the edge of the peek eyelet.The surgeon replaced with an ar-2324bcc, with lot# 11772467.No fragments were left in the patient from the broken eyelet piece and the surgery was completed successfully with no adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.75MM BC KNOTLESS SWIVELOCK
Common Device Name
BIO SOFT TISSUE FIXATN FASTNR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18300034
Report Number1220246-2023-09332
Device Sequence Number17932256
Product Code MAI
UDI-Device Identifier10888867351285
UDI-Public10888867351285
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K193503
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Model Number4.75MM BC KNOTLESS SWIVELOCK
Device Catalogue NumberAR-2324KBCC
Device Lot Number10984717
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2021
Initial Report FDA Received Date12/11/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-