| Model Number |
4.75MM BC KNOTLESS SWIVELOCK |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation; no pictures were provided.The most likely cause for the reported failure is user error.Per dfu-0087-11.G.Precautions.5.Pushlock and swivelock suture anchor only: insert the driver into the bone socket until the anchor body makes contact with the bone.Preview and adjust suture tension, if necessary.Tension will not increase during the final advancement of the anchor body.
|
| |
|
Event or Problem Description
|
|
It was reported that a biocomposite knotless swivelock, ar-2324kbcc, was being used in a procedure.When putting the anchor into the punched bone tunnel, the two fibertapes that were in the eyelet, ripped through the edge of the peek eyelet.The surgeon replaced with an ar-2324bcc, with lot# 11772467.No fragments were left in the patient from the broken eyelet piece and the surgery was completed successfully with no adverse effect to the patient.
|
| |
|