| Catalog Number |
LXMC17 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the device became discontinuous.Implanted on (b)(6) 2017.Patient hasn't had explant date scheduled.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 12/11/2023 b3: only event year known: 2023 attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Lot number was received and dhr is pending review.When the review is completed, a supplemental medwatch will be sent with a summary of the evaluation.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: what symptoms lead to the discovery of the discontinuous device? when did they begin? what was the date of the imaging which showed the discontinuous linx? if available, please share a copy of this imaging.Please send to: productcompliant1@its.Jnj.Com was the device initially effective in controlling reflux? were any events associated with the onset of symptoms (vomiting, retching, trauma, surgery)? did the patient undergo an mri since device implant? if so, when was the mri and what strength? did the patient have any other surgeries in the area? was any additional imaging performed since device implant? does the device appear to be in a continuous annular state in these images? we are interested in establishing a window when the device may have become discontinuous.Please share any additional images.What is the management plan? is device removal scheduled? is a replacement linx or fundoplication planned? when and if the explanation takes place can we ask that the procedure gets video recorded and the video shared? when and if the linx device is removed, may we ask that the device be returned for analysis? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Photo images were received.Any additional information obtained from the photo evaluation will be included as part of the product investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 3/27/2024.Photo analysis: two images of the device in vivo were reviewed by a medical safety officer.As per medical safety officer: "the fluoroscopic image demonstrates a discontinuous linx device.The photograph shows a device that has been removed." the mechanism/cause of failure is unknown as a hands-on analysis of the device is necessary to determine the cause of failure.No further investigation can be completed at this point.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 1/9/2024.Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.However, if the product is received at a later date, the investigation will be updated as applicable.A manufacturing record evaluation was performed for the finished device 12683 number, and no non-conformances related to the malfunction were identified.The device is affected by 2018 linx recall product bounding.
|
| |
|
Search Alerts/Recalls
|