| Model Number |
VTICM5_13.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/18/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2mm vticm5_13.2, -9.0/2.0/069 (sphere/cylinder/axis), implantable collamer lens into the patient's right eye (od).Excessive vault was observed.Lens remains implanted.Additional information has been requested but none has been forthcoming.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - work order search: no similar complaint type events were reported for units within the same lot.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B5: the reporter indicated that the surgeon implanted a 13.2mm vticm5_13.2, -9.0/2.0/069 (sphere/cylinder/axis), implantable collamer lens into the patient's right eye (od) on (b)(6) 2023.Lens was explanted and replaced on (b)(6) 2023 in a separate surgery with a shorter length lens due to excessive vault.Problem was resolved.Patient is satisfied.Cause of event was reported as other; unpredictable icl size calculation.Claim#: (b)(4).
|
| |
|