• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR SMOOTH ROUND MODERATE PLUS PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR SMOOTH ROUND MODERATE PLUS PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Catalog Number 3502200
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 11/10/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: deflation manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient underwent a breast augmentation revision with saline mentor smooth round moderate plus profile 200cc and experienced deflation on the left side post-operatively, which was confirmed during a physical exam.As a result, the patient underwent removal and replacement with mentor gel breast implants on (b)(6), 2023.
 
Additional Manufacturer Narrative
On jan 10, 2024, the mentor failure analysis lab received a device for evaluation with lot 9719701 (serial number (b)(6)).On jan 19, 2024, it was confirmed that this was the impacted product for this complaint.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Device evaluation summary: the product was returned to mentor for evaluation.Mentor then conducted visual inspection and microscopic examination of the returned device.Visual analysis of the returned sample determined that a tear was found on the posterior view extending to the anterior view of the sm mpps dv 200cc breast implant, measuring approximately 3.8 cm.Microscopic examination was performed on the edges of the rupture, and parallel striations were found in the whole area of the tear.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.Based on the information currently available, microscopic examination of the returned product indicates that the implant could have being damaged during or subsequent to implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps or scissors to contact the device during the implantation or other surgical procedures.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SMOOTH ROUND MODERATE PLUS PROFILE
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key18302482
Report Number1645337-2023-14547
Device Sequence Number15199590
Product Code FWM
UDI-Device Identifier00081317001355
UDI-Public00081317001355
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/12/2026
Device Catalogue Number3502200
Device Lot Number9719701
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/22/2023
Supplement Date Received by Manufacturer01/10/2024
Initial Report FDA Received Date12/11/2023
Supplement Report FDA Received Date02/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexFemale
-
-