U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV MEDICAL TECHNOLOGIES CHAMELEON; CATHETER, CONTINUOUS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV MEDICAL TECHNOLOGIES CHAMELEON; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number CH12-40-75US
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, when the device was pulled out of the packaging, the catheter shaft had split into two pieces.There was no obvious damage to the packaging and/or the catheter beforehand.The device's box came in intact and the sterile barrier was intact.Besides the reported issue, there was no other visible defects/damages found on the product.There was no leak and no luer adapter issue.There was no difficulty when pulling the device out of the packaging.There was no excessive force used on the device, and the detachment was seemingly spontaneous.The catheter was not repaired.There was no cleaning agent used on the device.To resolve the issue, a new product was provided to the customer from the trunk stock.There was no patient involvement.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Functional testing found that the cannula was found to be kinked, bent, or otherwise deformed.It was reported that the device has been detached.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHAMELEON
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
AV MEDICAL TECHNOLOGIES
21 habarzel street
tel-aviv 67770 16
IS  6777016
Manufacturer (Section G)
AV MEDICAL TECHNOLOGIES
21 habarzel street
tel-aviv 67770 16
IS   6777016
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18302504
Report Number3011144059-2023-00010
Device Sequence Number16950145
Product Code KRA
UDI-Device Identifier10884521812321
UDI-Public10884521812321
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K170635
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/17/2024
Device Model NumberCH12-40-75US
Device Catalogue NumberCH12-40-75US
Device Lot Number22203025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2023
Supplement Date Received by Manufacturer06/18/2024
Initial Report FDA Received Date12/11/2023
Supplement Report FDA Received Date06/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-