| Catalog Number |
07248563190 |
| Medical Device Problem Codes |
False Positive Result (1227); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
We received information that a customer observed discrepant results with 1 patient sample using the cobas® egfr mutation test v2.The alleged sample initially generated a "mutation detected" result for exon20ins.The same sample was retested on the same assay and generated a "mutation detected" result for exon20ins.The same sample was also retested using next generation sequencing (ngs) and was negative for the ex20ins.
|
| |
|
Additional Manufacturer Narrative
|
|
Roche received complaints from customers reporting the generation of false mutation detected results for the exon 20 insertion (ex20ins) mutation when using the cobas® egfr mutation test v2.During in-house testing using customer-provided ffpet samples, an ex20ins false mutation detected result was reproduced for one out of 8 ffpet samples, which was processed following the validated sample preparation method from the instructions for use.Although the majority of cases reported were from users using ffpet samples, the generation of false mutation detected ex20ins results with plasma specimens was reported in one case.Consignees have been notified of the issue with instruction to follow the instructions for use for sample input requirements.Additionally, if an ex20ins mutation detected result is generated with the cobas® egfr mutation test v2, customers must confirm the result with another method (e.G., sequencing or other pcr-based tests).
|
| |
|
Additional Manufacturer Narrative
|
|
Further review of the data revealed that the repeat result for the sample was mutation not detected for exon20ins.Based on the data that was provided, the sample generated results near the limit of detection of the assay.According to the method sheet, the detection of a mutation is dependent on the number of copies present in the sample and may be affected by sample integrity, the amount of isolated dna, and the presence of interfering substances.Medwatch fields h7 and h9 were updated.
|
| |
|
Search Alerts/Recalls
|