| Model Number |
HYF-V |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found the control unit was discolored.The light guide connector had a stain.The video connector had a stain.The charged coupled device unit was broken and no video output.The universal cord was crushed.There was a trace of fluid invasion in the device.The light guide lens had a stain.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus, the hysterovideoscope scope was unrecognized and the forceps were missing.The issue was found during receipt inspection.The device was returned for evaluation.During the device evaluation, the following reportable malfunction was found: the forceps channel port was not secured.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause of the loose forceps mouth issue could not be determined, however, the issue was likely the due to tightening a forceps/irrigation plug with excessive torque when attaching it to the lure mouthpiece (channel port), resulting in deformation of the forceps mouth.The event may be detected by following the instructions for use section below: chapter 3 preparation and inspection_ 3.3 inspection of the endoscope_ inspection of the endoscope.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|