MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTERLOCK; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTERLOCK; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 83779
Medical Device Problem Codes Stretched (1601); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr.: the device was returned for analysis.Visual inspection revealed only the main coil was returned for analysis.Additionally, it was observed that the main coil was stretched, detached, and bent at the coil arm section.The pouch was returned, and it was observed that the pouch information matched with the complaint information.No additional damages were observed.Functional inspection could not be performed due to only the main coil was returned for analysis.
 
Event or Problem Description
Reportable based on device analysis completed on (b)(6) 2023.It was reported that the coil could not be advanced in catheter and was stretched.A 6mm x 20cm f-idc 2d interlock was selected for use.During the procedure, the microcatheter could not be advanced after repeated adjustment.After the coil was withdrawn, it was found that the coil was stretched.The procedure was completed with a different device.No patient complications were reported and the patient was stable.However, device analysis revealed the detached main coil.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18307174
Report Number2124215-2023-67134
Device Sequence Number12458409
Product Code KRD
UDI-Device Identifier08714729765004
UDI-Public08714729765004
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/21/2026
Device Model Number83779
Device Catalogue Number83779
Device Lot Number0031091855
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/20/2023
Initial Report FDA Received Date12/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/21/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age71 YR
Patient SexMale
Patient Weight70 KG
-
-