Catalog Number S-60-150-120-P6
Medical Device Problem Codes Difficult to Remove (1528); Material Separation (1562); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/20/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the procedure was to treat a lesion in left superficial femoral artery.(sfa).A supera stent was implanted at the target lesion without issue.Then a 6.0x150mm supera stent was advanced distal of the previously implanted stent and the stent was implanted.When retracting the delivery system to be removed, resistance was noted with the thumbslide, and while pulling back the delivery system a break occurred.Under fluoroscopy, the distal end of the delivery system (near the nose cone) was observed to be separated; however, remained on the guide wire.A guide catheter was used to entrap the separated portion and remove it from the patient.Both supera remained implanted with no issues.There was no adverse patient sequela.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported tip separation was able to be confirmed.The reported mechanical jam was unable to be replicated in a testing environment due to the condition of the returned device.The reported difficult to remove was unable to be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that during removal interaction with the anatomy and/or the previously implanted supera stent resulted in the reported mechanical jam (resistance was noted with the thumbslide) and the reported difficult to remove.Interaction/manipulation of the compromised device ultimately resulted in the reported tip separation/noted inner member and tip jacket separations.The treatment appears to be related to the operational context of the procedure as reportedly a guide catheter was used to entrap the separated portion and remove the separation from the patient.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18307622
Report Number2024168-2023-13812
Device Sequence Number12413349
Product Code NIP
UDI-Device Identifier08717648211843
UDI-Public08717648211843
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P120020
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Catalogue NumberS-60-150-120-P6
Device Lot Number3051561
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/20/2023
Supplement Date Received by Manufacturer01/11/2024
Initial Report FDA Received Date12/11/2023
Supplement Report FDA Received Date01/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SUPERA STENT.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-