• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARGON MEDICAL DEVICES CLOSUREFAST MICRO INTRODUCER SHEATH SET 7FR/CH (2.3MM) X 7CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARGON MEDICAL DEVICES CLOSUREFAST MICRO INTRODUCER SHEATH SET 7FR/CH (2.3MM) X 7CM Back to Search Results
Model Number MIS-7F07
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
According to the product experience report, sample is not available for return.Argon is investigating this reported event and a follow-up report will be submitted upon investigation completion.
 
Event or Problem Description
Physician attempted to use a micro introducer kit during patient treatment.The lumen was flushed prior to use.The ifu (instruction for use) was followed during preparation, procedure, post-procedure.It was reported the valve stayed open when they pulled out the dilater and 018 wire before they put the catheter in and the patient bleed everywhere.A replacement mis-7f07 kit was used to complete procedure.No additional treatment required.No patient injury reported.
 
Additional Manufacturer Narrative
Corrected lot number 11493083.
 
Event or Problem Description
Physician attempted to use a micro introducer kit during patient treatment.The lumen was flushed prior to use.The ifu (instruction for use) was followed during preparation, procedure, post-procedure.It was reported the valve stayed open when they pulled out the dilater and 018 wire before they put the catheter in and the patient bleed everywhere.A replacement mis-7f07 kit was used to complete procedure.No additional treatment required.No patient injury reported.
 
Event or Problem Description
Physician attempted to use a micro introducer kit during patient treatment.The lumen was flushed prior to use.The ifu (instruction for use) was followed during preparation, procedure, post-procedure.It was reported the valve stayed open when they pulled out the dilater and 018 wire before they put the catheter in and the patient bleed everywhere.A replacement mis-7f07 kit was used to complete procedure.No additional treatment required.No patient injury reported.
 
Additional Manufacturer Narrative
A review of the dhr and inspection records was conducted, and no deviations or non-conformances were found.No samples or images were provided to argon by the customer.Since this is a known issue, the complaint was confirmed.Capa ca-00058 has been initiated to address the introducer leakage issue.The capa will document the root cause identified and the corrective action that was taken to address the issue.
 
Additional Manufacturer Narrative
Udi related data quality updates only.Corrected information provided in d.4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLOSUREFAST MICRO INTRODUCER SHEATH SET 7FR/CH (2.3MM) X 7CM
Common Device Name
CLOSUREFAST MICRO INTRODUCER SHEATH SET
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX 75751
Manufacturer Contact
elnaz rahman
1445 flat creek rd
athens, TX 75751
9036759321
MDR Report Key18307993
Report Number0001625425-2023-01261
Device Sequence Number6783756
Product Code DRE
UDI-Device Identifier10884521140295
UDI-Public(01)10884521140295(17)280815(10)11493083
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K780126
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 07/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMIS-7F07
Device Lot Number11493083
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/27/2023
Supplement Date Received by Manufacturer11/27/2023
11/27/2023
07/08/2024
Initial Report FDA Received Date12/11/2023
Supplement Report FDA Received Date12/11/2023
05/17/2024
07/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-