| Model Number |
MIS-7F07 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Physician attempted to use a micro introducer kit during patient treatment.The lumen was flushed prior to use.The ifu (instruction for use) was followed during preparation, procedure, post-procedure.It was reported the valve stayed open when they pulled out the dilater and 018 wire before they put the catheter in and the patient bleed everywhere.A replacement mis-7f07 kit was used to complete procedure.No additional treatment required.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
According to the product experience report, sample is not available for return.Argon is investigating this reported event and a follow-up report will be submitted upon investigation completion.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the dhr and inspection records was conducted, and no deviations or non-conformances were found.No samples or images were provided to argon by the customer.Since this is a known issue, the complaint was confirmed.Capa (b)(4) has been initiated to address the introducer leakage issue.The capa will document the root cause identified and the corrective action that was taken to address the issue.
|
| |
|
Event or Problem Description
|
|
Physician attempted to use a micro introducer kit during patient treatment.The lumen was flushed prior to use.The ifu (instruction for use) was followed during preparation, procedure, post-procedure.It was reported the valve stayed open when they pulled out the dilater and 018 wire before they put the catheter in and the patient bleed everywhere.A replacement mis-7f07 kit was used to complete procedure.No additional treatment required.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi related data quality updates only.Corrected information provided in d.4.
|
| |
|