| Model Number |
89-10501 |
| Medical Device Problem Codes |
Material Puncture/Hole (1504); Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A complaint was received on 11/20/2023 hole in back table cover.The actual sample has not been returned to deroyal for evaluation at this time.This investigation is ongoing at this time.No further information is available at this time.We will provide follow-up report if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
Hole in back table cover.
|
| |
|
Additional Manufacturer Narrative
|
|
A complaint was received on 11/20/2023 hole in back table cover.The sample has been returned to deroyal for evaluation.The root cause was determined to be from the placement of the yankauer within the tray.Corrective action: quality control specialist contacted tray engineering to evaluate the tray for a placement change on the yankauer within the tray.Engineering change order number 60184 has been created for a placement change for the yankauer.This investigation is complete at this time.No further information is available at this time.We will provide follow-up report if additional information becomes available.
|
| |
|