ZIMMER BIOMET, INC. VNGD SSK 360 R FEM; PROSTHESIS KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that surgeon found the sterile tray was cracked.Another device was used to complete the surgery.Attempts to obtain additional information have been made; however, no more is available.
 
Additional Manufacturer Narrative
(b)(4).G2: japan.H3: customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned device shows the sterile packaging is damaged with corresponding damage to the inside of the carton.This indicates the device shifted during transit contributing to the reported damage.Sterility has been breached.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.The root cause of the reported event can be attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VNGD SSK 360 R FEM
Common Device Name
PROSTHESIS KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18310118
Report Number0001825034-2023-02957
Device Sequence Number17803145
Product Code JWH
UDI-Device Identifier00880304481978
UDI-Public(01)00880304481978(17)260525(10)3800168
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K093293
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/25/2026
Device Model NumberN/A
Device Catalogue Number185266
Device Lot Number3800168
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2023
Supplement Date Received by Manufacturer01/11/2024
Initial Report FDA Received Date12/12/2023
Supplement Report FDA Received Date01/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/25/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-