(b)(4).Date sent: 12/12/2023.D4 batch # 332c05.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.Photo images were received.Any additional information obtained from the photo evaluation will be included as part of the product investigation.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
(b)(4) date sent: 1/12/2024 d4: batch #297c63 investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the tx60g device was received along with tree reloads (b, c and d) present.The reloads b and c were received fully fired damaged as the rear cap was broken off and no returned and with pin and rear cap missing.The reload d was received fully loaded with staples and damaged as the rear cap was broken off and no returned as well and with the pin missing.The device was tested for functionality with the returned reload and it fired and formed all the staples as intended.The device fired without any difficulties, the staple line was complete, and the staples meet the staple form release criteria.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as no malformed staples were noted.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: when positioning the stapler on the application site, ensure that no obstructions such as clips, stents, guide wires, etc., are within the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.It is possible that the reload was not loaded correctly.To reload the device the tracks on each side of the reload should be used as guides to align the reload within the jaws of the instrument.When the reload is properly aligned, push the reload into the instrument until it is fully seated, and a click is heard.If the instructions are not following the reload may became damage.A manufacturing record evaluation was performed for the finished device batch number 297c63, and no non-conformances were identified.
|