• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH INC PEN NEEDLES FG; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH INC PEN NEEDLES FG; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number PNDL, 5BV TVH TP 31G 6MM100CT 30/CS
Medical Device Problem Codes Material Twisted/Bent (2981); No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/21/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal report reference number: (b)(4).Additional report reference number: (b)(4).Pen needles were not returned for evaluation.Note: manufacturer contacted customer in a follow-up call on (b)(6) 2023 to ensure the replacement products resolved the initial concern, customer stated they are comfortable with the replacement products.
 
Event or Problem Description
Consumer reported complaint for the trueplus pen needles.Customer called and stated that she has two boxes of the 31g pen needles and with both boxes, she noticed that some of the needles are bent and some of the needles are not aspirating correctly; additional report reference number (b)(4).The package had not been open or damaged when received by the customer.The customer has been using the product for about one month.Customer is using compatible product and the pen needle is aligned properly.At the time of the call the customer felt well and did not report any symptoms.Customer did not claim to be injured while using the pen needles and no medical intervention related to the use of the product was reported.
 
Additional Manufacturer Narrative
Sections with additional information as of 17-apr-2024: d9: device available for evaluation.H3: device evaluated by manufacturer.H6: updated fda¿s type, findings and conclusions codes.H10: pen needle was returned - customer returned 1 used pen needle without the protective cover and cover cap; unable to ship to the manufacturer due to needle exposed.Return product scrapped.Complaint was forwarded to supplier quality and internal evaluation was performed by the manufacturer using pen needles from the same lot.No abnormalities observed with retention samples.Most likely underlying root cause: mlc-009: use error caused or contributed to event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEN NEEDLES FG
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key18311055
Report Number1000113657-2023-00603
Device Sequence Number12331825
Product Code FMI
UDI-Device Identifier00021292010782
UDI-Public(01)00021292010782
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K180720
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberPNDL, 5BV TVH TP 31G 6MM100CT 30/CS
Device Lot Number2H621
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Distributor Facility Aware Date11/21/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/21/2023
Supplement Date Received by Manufacturer03/25/2024
Initial Report FDA Received Date12/12/2023
Supplement Report FDA Received Date04/17/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-