Catalog Number X21417
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The varady phleb lge hk/spatu 7 in de extractor (x21417) was not returned for evaluation, as the customer stated the product was disposed of.Lot number was provided, device history record (dhr) was reviewed, and no abnormalities related to the reported issue were identified.Customer provided a photo of the instrument.Per the provided image, it was determined that the tip breakage may have resulted from environmental damage or rough handling.However, a definitive root cause cannot be determined without return of the actual product for evaluation.No manufacturing, workmanship or material deficiency has been identified.
 
Event or Problem Description
A facility reported that the tip of the varady phleb lge hk/spatu 7 in de extractor (x21417) broke off in a patient while pulling on a vein, during vascular surgery.An x-ray was taken to retrieve the broken part.It was reported that the patient was in good condition.Another device was available for use.There was a delay in surgery of approximately 20-30 minutes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VARADY PHLEB LGE HK/SPATU 7 IN DE EXTRACTOR
Common Device Name
JARIT
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key18312368
Report Number3014334038-2023-00185
Device Sequence Number20512914
Product Code GAI
UDI-Device Identifier10381780373582
UDI-Public10381780373582
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2023
Device Catalogue NumberX21417
Device Lot NumberAC2306
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/15/2023
Initial Report FDA Received Date12/12/2023
Date Device Manufactured06/01/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-