• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD SAF-T-INTIMA W/PRN YEL 24GA X .75IN; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD SAF-T-INTIMA W/PRN YEL 24GA X .75IN; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383312
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/14/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd saf-t-intima w/prn yel 24ga x.75in tubing was defective / damaged the following information was provided by the initial reporter; "1.Specific operational use: infected hepatology ward on november 13, according to medical advice to the patient retained peripheral venous indwelling needle, the remaining november 14, 9:00 a.M.Found that the indwelling needle clamp buckle cracked seepage.2.Adverse event situation: the clip buckle of the indwelling needle cracked and leaked fluid.3.Impact on the victim: the leaking fluid soaked the patient's gown, causing no harm to the patient.4.Treatment measures taken and outcome: immediately replace the indwelling needle with a new one after discovery and report the adverse event".
 
Additional Manufacturer Narrative
A device history record review was completed by our quality engineer team for provided material number: 383312 and lot number: 2089814.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.Should you again experience any problems with our product we would appreciate the opportunity to conduct a thorough analysis.There are quality controls currently in place to detect this type of defect during the production process.Further action has not been determined necessary at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
No additional information received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SAF-T-INTIMA W/PRN YEL 24GA X .75IN
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales UT
MX  
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18314648
Report Number9610847-2023-00348
Device Sequence Number6114942
Product Code FOZ
UDI-Device Identifier30382903833123
UDI-Public(01)30382903833123
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K923702
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date03/31/2026
Device Catalogue Number383312
Device Lot Number2089814
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2023
Supplement Date Received by Manufacturer02/22/2024
Initial Report FDA Received Date12/12/2023
Supplement Report FDA Received Date02/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-