• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES HEMOSPHERE; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES HEMOSPHERE; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number HEMCSM10
Medical Device Problem Codes Incorrect Measurement (1383); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/05/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product has not been returned for evaluation.When it arrives and the product evaluation has been completed, a supplemental report will be submitted.The device history record review was completed and all inspections passed with no non-conformances.Complaint histories for all reported events are reviewed against trending control limits, and any excursions above the control limits are assessed and documented as part of this monthly review.H3 other text : awaiting device return.
 
Event or Problem Description
It was reported that the hemcsm10 blood pressure reading is inconsistent.The blood pressure readings were sporadic in that the systolic would become low then increase quickly without any interventions.It continued to happen so staff checked placement of the cuff made sure no lines were kinked and the hrs was re-zeroed.No damage was noted.This was during use.There was no inappropriate treatment administered and patient demographics were available.
 
Additional Manufacturer Narrative
A supplemental mdr is being submitted due to product evaluation findings.Sections g6, h2, h6: type of investigation, investigation findings, and investigation conclusions have been updated.One hemcsm10 was received for product evaluation.The suspect hemcsm10 was connected to a known working hem1 hot mock-up system.The hemcsm10 caused no errors.It was also able to obtain normal blood pressure readings and waveform.The hemcsm10 was ran for half an hour without any errors.No damage was found.There was no defect.
 
Additional Manufacturer Narrative
Udi related data quality updates only the following are additional product codes for this device: dqe catheter, oximeter, fiber-optic, qaq adjunctive predictive cardiovascular indicator, mud oximeter, tissue saturation, dsb plethysmograph, impedance, qms adjunctive open loop fluid therapy recommender, fll thermometer, electronic, clinical, dxn system, measurement, blood-pressure, non-invasive.The following is corrected data, as the original identification information sent did not match the data submitted to the global unique device identification database (gudid): brand name: hemosphere.D2a.Common device name: computer, diagnostic, programmable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOSPHERE
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
one edwdards way
irvine CA 92614
Manufacturer Contact
katheryn richards
1 edwards way
irvine, CA 92614
9492017706
MDR Report Key18314993
Report Number2015691-2023-18324
Device Sequence Number17838603
Product Code DQK
UDI-Device Identifier00690103202762
UDI-Public(01)00690103202762(11)210304
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K203687
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/05/2026
Device Model NumberHEMCSM10
Device Lot Number16800882
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/05/2023
Supplement Date Received by Manufacturer01/29/2024
07/09/2024
Initial Report FDA Received Date12/12/2023
Supplement Report FDA Received Date02/13/2024
07/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/05/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-