The product has not been returned for evaluation.When it arrives and the product evaluation has been completed, a supplemental report will be submitted.The device history record review was completed and all inspections passed with no non-conformances.Complaint histories for all reported events are reviewed against trending control limits, and any excursions above the control limits are assessed and documented as part of this monthly review.H3 other text : awaiting device return.
|
|
A supplemental mdr is being submitted due to product evaluation findings.Sections g6, h2, h6: type of investigation, investigation findings, and investigation conclusions have been updated.One hemcsm10 was received for product evaluation.The suspect hemcsm10 was connected to a known working hem1 hot mock-up system.The hemcsm10 caused no errors.It was also able to obtain normal blood pressure readings and waveform.The hemcsm10 was ran for half an hour without any errors.No damage was found.There was no defect.
|
|
Udi related data quality updates only the following are additional product codes for this device: dqe catheter, oximeter, fiber-optic, qaq adjunctive predictive cardiovascular indicator, mud oximeter, tissue saturation, dsb plethysmograph, impedance, qms adjunctive open loop fluid therapy recommender, fll thermometer, electronic, clinical, dxn system, measurement, blood-pressure, non-invasive.The following is corrected data, as the original identification information sent did not match the data submitted to the global unique device identification database (gudid): brand name: hemosphere.D2a.Common device name: computer, diagnostic, programmable.
|