• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE 90-S ACCELERATOR; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE 90-S ACCELERATOR; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 0279351400
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/13/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that during procedure material detached from the probe.The pieces were potentially unable to be retrieved.
 
Additional Manufacturer Narrative
Alleged failure: during the surgery material detached from the probe.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be excessive force used when inserting or removing the probe, a probe inserted into an obstructed passageway, or any forceful impact to the tip of the probe with rigid objects.The probe is used as a tool for a mechanical displacement of tissue or bone, or to pry or scrub.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that during procedure material detached from the probe.The pieces were potentially unable to be retrieved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
90-S ACCELERATOR
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key18315365
Report Number0002936485-2023-01235
Device Sequence Number6278195
Product Code GEI
UDI-Device Identifier07613327057041
UDI-Public07613327057041
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K160050
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0279351400
Device Lot Number23083AE2
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/15/2023
Supplement Date Received by Manufacturer11/15/2023
Initial Report FDA Received Date12/12/2023
Supplement Report FDA Received Date06/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/24/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-