| Catalog Number |
0279351400 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that during procedure material detached from the probe.The pieces were potentially unable to be retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: during the surgery material detached from the probe.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be excessive force used when inserting or removing the probe, a probe inserted into an obstructed passageway, or any forceful impact to the tip of the probe with rigid objects.The probe is used as a tool for a mechanical displacement of tissue or bone, or to pry or scrub.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that during procedure material detached from the probe.The pieces were potentially unable to be retrieved.
|
| |
|