• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICMO 12.1
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/11/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A4-a6: unk.Claim #(b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted a 12.1mm vticmo 12.1 implantable collamer lens of diopter -14.5/3.0/093 (sphere/cylinder/axis) into the patient's right eye (od) on (b)(6) 2023.On (b)(6) 2023 the lens was exchanged for a longer length lens due to low vault with rotation.The problem was resolved.The cause of the event was reported as unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key18315519
Report Number2023826-2023-05603
Device Sequence Number6267782
Product Code QCB
UDI-Device Identifier00840311319510
UDI-Public00840311319510
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Model NumberVTICMO 12.1
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2023
Initial Report FDA Received Date12/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL: SFC-45 - LOT# UNK.; FOAM TIP PLUNGER - LOT# UNK.; INJECTOR MODEL: MSI-PF - LOT# UNK.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age26 YR
Patient SexFemale
-
-