| Model Number |
WA4KL100 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned and evaluated, and the customer¿s allegation was not confirmed.Additional findings were decrease of light transmission due to broken fibers and scratches on outer tube.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus that the telescope "ultra", 10 mm, 0° had no blurry image, only a few fibers were broken.There was no patient harm associated with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event could not be determined although it can be presumed it was due to excessive force by the customer.Olympus will continue to monitor field performance for this device.
|
| |
|