• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. 32MM I.D. SIZE HH NEUTRAL LINER USE WITH 50MM O.D. SIZE HH SHELL; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. 32MM I.D. SIZE HH NEUTRAL LINER USE WITH 50MM O.D. SIZE HH SHELL; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 11/20/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2023 - 03530, 0001822565 - 2023 - 03532.D10: cat# 00811400110 lot# unk femoral stem 12/14 neck taper ext.Offset size 1 130 mm stem length.G2: foreign: united kingdom.The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient underwent a total hip replacement.Subsequently, the patient had a fall, landing on the hip that was replaced, causing the dislocation.The patient underwent a revision procedure approximately 1 month post implantation.Perioperatively, the surgeon decided to exchange the shell due to shell malposition and the liner for an elevated liner.On removal of the femoral head, the stem was removed and exchanged for the same offset and size.It was reported that no further information is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Lot identification is necessary for review of device history records, lot identification was not provided.Medical records were not provided.It is unknown if the shell was placed incorrectly, or was placed that way dependent on patient anatomy.While the cup position could be a contributing factor to the dislocation, a definitive root cause cannot be identified.Additionally, the reason for the fall is unknown.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further information at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected}.Updated: d4; d9; g3; h2.
 
Event or Problem Description
No further information at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected}.Updated: d4; d5; g3; h2; h3; h4; h6: h10.Visual examination of the returned product identified liner was returned seated in the shell.Shell has bio debris embedded in the outer spherical surface.Shell rim face is discolored, gouged, and indented.Liner shows one anti rotation scallop missing, screw hole, and gouges in the spherical surface.Liner has rim wear consistent to the reported dislocation.No other damage was noted.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Root cause unchanged.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further information at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
32MM I.D. SIZE HH NEUTRAL LINER USE WITH 50MM O.D. SIZE HH SHELL
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18317686
Report Number0001822565-2023-03531
Device Sequence Number6152171
Product Code JDI
UDI-Device Identifier00889024149106
UDI-Public(01)00889024149106(17)280310(10)65843426
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K200823
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number00875100932
Device Lot Number65843426
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2023
Supplement Date Received by Manufacturer03/25/2024
03/20/2025
09/04/2025
Initial Report FDA Received Date12/13/2023
Supplement Report FDA Received Date03/28/2024
04/01/2025
09/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/12/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-