| Catalog Number |
04810716190 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The analyzer serial number is (b)(6).The calibration and qc recovery data provided was acceptable.The alarm trace showed abnormal probe sucking alarms and several water reservoir level too low and bath water sensor alarms.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable crej2 creatinine jaffé gen.2 results for 1 patient sample on a cobas 4000 c (311) stand alone system.The initial crej2 result was 3.07 mg/dl.An aliquot of the sample was repeated and the result was 0.98 mg/dl.On 19-nov-2023, the sample was repeated again and the result was 3.22 mg/dl.The patient's doctor deemed the first repeat result as correct.No questionable results were reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the calibration and qc data, a general reagent issue could be excluded.The investigation did not identify a product problem.The root cause of the event was found to be consistent with pre-analytical sample handling issues.
|
| |
|