• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. ETHICON SECURESTRAP 5MM ABSORBABLE STRAP FIXATION DEVICE; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. ETHICON SECURESTRAP 5MM ABSORBABLE STRAP FIXATION DEVICE; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number STRAP25R
Medical Device Problem Code Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a laparoscopic hernia repair procedure on an unknown date in 2023 and absorbable straps were used to fix a competitor's mesh.The device fires a fixation clip, then misfires/ doesn't fire.The clips/strap did not fix to the abdominal wall.Counter pressure was applied.Opened 2nd device and sutured with prolene.The surgery was delayed due to the reported event, but successfully completed.There were no adverse patient consequences reported.Additional information has been requested.
 
Additional Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.How long was the procedure delayed? were there any adverse patient consequences? product lot number? attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.H6 component code: g07002 ¿ device not returned.
 
Additional Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 analysis summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the strap25r device was returned with no apparent damage in the external components.In an attempt to replicate the reported incident, the provided device was activated manually.Nineteen(19) straps were delivered properly.The device trigger was locked as a normal process because the lock-out indicator turned 100% orange color.After testing the device was disassembled to conduct a deep inspection and no damages were found on internal instrument components.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what may have caused the reported incident.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.A review of the batch manufacturing records was conducted, and no related non-conformances were identified.
 
Additional Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected information: d4 udi.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHICON SECURESTRAP 5MM ABSORBABLE STRAP FIXATION DEVICE
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key18318980
Report Number2210968-2023-09731
Device Sequence Number7358768
Product Code GDW
UDI-Device Identifier10705031470552
UDI-Public(01)10705031470552(11)220912(17)240831(10)SKMHAS
Combination Product (Y/N)N
Initial Reporter CountryWA
PMA/510(K) Number
K093845
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Catalogue NumberSTRAP25R
Device Lot NumberSKMHAS
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/21/2023
Supplement Date Received by Manufacturer05/23/2024
07/16/2024
Initial Report FDA Received Date12/13/2023
Supplement Report FDA Received Date06/18/2024
07/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-