|
One tubing set without lot number identification was returned for evaluation.The tubing was functionally tested on a scope and no issues were noted; the reported event could not be duplicated.The endogator hybrid irrigation tubing instructions for use states, "place the printed section of the irrigation tubing, where the text "place in pump head." is printed, in the pump head and close.(note: ensure that the text "place in pump head." is centered, aligned and carefully pulled taut before closing the pump head.)" for endogator hybrid irrigation tubing models with connectors defining the tube segment placed within the pump head, the ifu states, "open the pump head, insert the pump tube of the irrigation tubing and close the pump head.(note: ensure that the pump tubing between the tubing connectors is centered, aligned, and carefully pulled taut before closing the pump head.)" the ifu further states, "connect the air/water connector of the endogator hybrid irrigation tubing to the air/water port of the gi endoscope with a firm pushing motion.Prime the tubing and the auxiliary water channel prior to use by activating the foot pedal; flush water through the entire gi endoscope.Turn on the air processor or co2.Prime the channel and test for air and water prior to insertion of gi endoscope into patient.If the water pressure is low, then ensure that the water bottle is tightly closed." no additional issues have been reported.
|