| Catalog Number |
DLF180733 |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section d2b ¿ procode: krd/hcg.Section e1.Initial reporter phone: (b)(6).The device was discarded; therefore, no further investigation can be performed.A manufacturing record evaluation was performed for the finished device 30768264 number, and no non-conformances related to the malfunction were identified.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The company is seeking this information through the event investigation.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
As reported by the field, this was a a coil embolization of the right internal iliac artery prior to evar, using 5fr diagnostic catheter and coiling support.The above system was advanced to the lesion.After implanted 2 of the j&j coils, a deltafill18 7mm x 33cm coil (dlf180733, 30768264) was used as a third coil.Attempted to detach but could not.Attempted again but could not detach the coil.The same cable was used first to third coil.The cable was replaced with another new one, but the issue was not resolved.Replaced the coil (dlf180733) with another new one, which did detach.The procedure was completed with total 5 coils implanted.There was no negative impact to the patient.A continuous flush was done.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Complaint conclusion: as reported by the field, this was a a coil embolization of the right internal iliac artery prior to evar, using 5fr diagnostic catheter and coiling support.The above system was advanced to the lesion.After implanted 2 of the j&j coils, a deltafill18 7mm x 33cm coil (dlf180733, 30768264) was used as a third coil.Attempted to detach but could not.Attempted again but could not detach the coil.The same cable was used first to third coil.The cable was replaced with another new one, but the issue was not resolved.Replaced the coil (dlf180733) with another new one, which did detach.The procedure was completed with total 5 coils implanted.There was no negative impact to the patient.A continuous flush was done.The device was discarded; therefore, no further investigation can be performed.A manufacturing record evaluation was performed for the finished device 30768264 number, and no non-conformances related to the malfunction were identified.Failure to detach is a known potential issue associated with the use of this device.The instructions for use (ifu) contains several precautions related to this issue and includes instructions for troubleshooting the situation should it be encountered during use.With the information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the manufacturing record evaluation, there is no indication that the event is related to the device manufacturing process.Multiple attempts to obtain additional information were unsuccessful.If information is provided at a later date, the file will be reopened and processed accordingly.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.
|
| |
|
Search Alerts/Recalls
|
|