• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARGON MEDICAL DEVICES BIOPINCE ULTRA FULL CORE BIOPSY INSTRUMENT 18GA X 20CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARGON MEDICAL DEVICES BIOPINCE ULTRA FULL CORE BIOPSY INSTRUMENT 18GA X 20CM Back to Search Results
Model Number 360-2080-02
Medical Device Problem Code Failure to Obtain Sample (2533)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Ten incidents were reported under one product experience report.Argon reached out to the reporter for additional information.Argon did not receive any additional information or ten separate reports for the incidents.The 10 incidents are reported under (comp-(b)(4) ).According to the report, two out of ten samples will be returned for investigation.At this point, it is unclear which samples will get returned.Argon is investigating the reported events and a follow-up report will be submitted when new information becomes available and investigation is completed.
 
Event or Problem Description
Out of 16 biopsies, there were 10 incidents.5 times the gun failed after two or three uses, making vacuum punctures without bringing back pieces of tissue, which increases the attack on the prostate and therefore the risk of hemorrhage and infection and which required for once, 7 guns to make 12 carrots (on average 2) 6 times the procedure was followed by hospitalization or consumption of heavy treatment.- a night hospitalization at the (b)(6) hospital for the same situation of hemorrhage/catheterization with a week of hospitalization for a fragile grandfather - prostatitis with prescription of rocéphine at home - hospitalization in (b)(6) for severe prostatitis bordering on septic shock the same evening of the puncture, rocéphine amiklin, 4 days of hospitalization - a survey for acute urine retention - a hematoma responsible for dysuria, prescription of bactrim and alpha blocker - urethrorrhagia giving in to hyper hydration.(see below for notes/ additional information).
 
Additional Manufacturer Narrative
A review of the manufacturing and inspection records for this lot was conducted.No deviations or non-conformances were found.One unopened sample was returned from the customer for review.A visual inspection was performed on the returned product.Failure to obtain samples was due to the difference from the pincer tip to the window on line 2, which drifted from the settings called out in the procedure.The root cause was identified under capa ca-00042.Line 2's inspection parameters were reset to match the op.This change was then implemented into the software for other gauge sizes.Batches showed improvement based on sampling beef kidneys.
 
Event or Problem Description
Out of 16 biopsies, there were 10 incidents.5 times the gun failed after two or three uses, making vacuum punctures without bringing back pieces of tissue, which increases the attack on the prostate and therefore the risk of hemorrhage and infection and which required for once, 7 guns to make 12 carrots (on average 2) 6 times the procedure was followed by hospitalization or consumption of heavy treatment.A night hospitalization at (b)(6) hospital for the same situation of hemorrhage/catheterization with a week of hospitalization for a fragile grandfather.Prostatitis with prescription of rocéphine at home.Hospitalization in (b)(6) for severe prostatitis bordering on septic shock the same evening of the puncture, rocéphine amiklin, 4 days of hospitalization.A survey for acute urine retention.A hematoma responsible for dysuria, prescription of bactrim and alpha blocker.Urethrorrhagia giving in to hyper hydration.
 
Additional Manufacturer Narrative
Udi related data quality updates only corrected information provided in d.4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOPINCE ULTRA FULL CORE BIOPSY INSTRUMENT 18GA X 20CM
Common Device Name
BIOPINCE
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX 75751
Manufacturer Contact
elnaz rahman
1445 flat creek rd
athens, TX 75751
9036759321
MDR Report Key18322938
Report Number0001625425-2023-01274
Device Sequence Number7081315
Product Code KNW
UDI-Device Identifier00886333224043
UDI-Public(01)00886333224043(17)260328(10)11469345
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K904987
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/28/2026
Device Model Number360-2080-02
Device Lot Number11469345
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer12/01/2023
09/04/2024
Initial Report FDA Received Date12/13/2023
Supplement Report FDA Received Date02/28/2024
09/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
-
-