MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; RESPIRATORY HUMIDIFIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; RESPIRATORY HUMIDIFIER Back to Search Results
Model Number MR850ALU
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Excessive Heating (4030)
Health Effect - Clinical Code Full thickness (Third Degree) Burn (2696)
Type of Reportable Event Serious Injury
Event or Problem Description
A healthcare facility in spain reported via a fisher & paykel healthcare (f&p) field representative, that a neonate sustained a third degree burn to the nose when receiving humidification from a f&p healthcare mr850 respiratory humidifier via a f&p healthcare rt265 infant evaqua2 breathing circuit and a nasal mask manufacturered by a third party.It was also reported that the patient required additional medical treatment and surgery as a result of the injury.There were no further patient consequences reported.
 
Additional Manufacturer Narrative
(b)(4).The complaint mr850 respiratory humidifier is currently en route to fisher & paykel (f&p) healthcare for evaluation to determine the involvement of our product in the reported event.We will provide a follow-up report upon completion of our investigation.
 
Additional Manufacturer Narrative
(b)(4).Corected data: d4 - serial number and udi (unique device identifier).D9 - device returned to manufacturer - date of return.H6 - health effect - impact code.- medical device problem code.Method: the subject mr850 respiratory humidifier (sn (b)(6), the 900mr805 heater wire adaptor (lot 2101452494) and the 900mr869 temperature/flow probe (lot 140407) in use during the reported event were returned to the fisher & paykel (f&p) healthcare in new zealand where they were visually inspected and functionally tested.Our investigation is based on the information provided by the healthcare facility, f&p healthcare's evaluation of the returned devices, and our knowledge of the products.Results: visual inspection of the mr850 respiratory humidifier identified some minor damage to the enclosure of the mr850 respiratory humidifier.There were no defects or damage observed on the 900mr805 heater wire adaptor or the 900mr869 temperature/flow probe.The mr850 respiratory humidifier and the 900mr869 temperature/flow probe passed functional testing, including the alarm system check.Other than the enclosure damage, no faults or malfunctions were identified with either device.During the functional testing of the 900mr805 heater wire adaptor, an intermittent heater wire alarm occurred during the startup of the subject mr850 respiratory humidifier.Further testing of the 900mr805 heater wire adaptor confirmed an intermittent continuity between one of the heater wire connector pins and its wire.The intermittent connection of the 900mr805 heater wire adaptor would not cause or contribute to the reported elevated temperature on the mr850 respiratory humidifier.A faulty heater wire adaptor would generate an alarm condition on the mr850 respiratory humidifier, and the humidifier will immediately turn off power to the heater wires and heater plate.The healthcare facility did not confirm that any alarm conditions occurred at the time of the reported event.Conclusion: based on the information provided by the healthcare facility and the investigation we have undertaken we have been unable to determine the cause of the reported temperature fluctuation on the mr850 respiratory humidifier or the injury sustained by the patient.The subject mr850 respiratory humidifier and the 900mr869 temperature/flow probe were confirmed to be operating as intended and within specification.The mr850 respiratory humidifier is used to warm and humidify gases delivered to patients requiring mechanical ventilation, positive pressure breathing assistance or medical gases.The mr850 complies with iso8185:2007: "respiratory tract humidifiers for medical use - particular requirements for respiratory humidification systems." this standard specifies safety requirements to prevent the risk of thermal injury, including maximum enthalpy limits for humidified gas delivered to patients through respiratory humidification systems.The mr850 respiratory humidifier features an audible and visual alarm which alerts the user if the displayed temperature exceeds 41°c, or if the airway temperature exceeds 43°c and immediately disables the heater wire and the heater-plate.The mr850 respiratory humidifier user instruction states the following warning: ensure that both temperature probe sensors are correctly and securely fitted.Failure to do so may result in temperatures in excess of 41°c being delivered to the patient.The rt265 infant evaqua2 breathing circuit user instruction also states the following warnings: do not use heated wire breathing circuits without gas flow.If gas flow is interrupted, turn the humidifier off.Avoid prolonged contact of heated tubes with patient's skin.Check all connections are tight before use.
 
Event or Problem Description
A healthcare facility in spain reported via a fisher & paykel (f&p) healthcare field representative, that a patient sustained a facial burn to the nose when receiving humidified gases while using a non-f&p cpap nasal mask and a mr850 respiratory humidifier.It was further reported that the injury was a third degree burn and required medical treatment and surgery.Subsequent to that, the patient's condition has been reported as stable while residing at home.
 
Event or Problem Description
A healthcare facility in spain reported via a fisher & paykel (f&p) healthcare field representative, that a patient sustained a facial burn to the nose when receiving humidified gases while using a non-f&p cpap nasal mask and a mr850 respiratory humidifier.It was further reported that the injury was a third degree burn and required medical treatment and surgery.Subsequent to that, the patient's condition has been reported as stable while residing at home.
 
Additional Manufacturer Narrative
(b)(4).**udi related data quality updates only** corrections to reflect gudid: section d1: brand name updated to fisher & paykel healthcare.Section d2a: common device name updated to respiratory humidifier.Section d4: model and catalog # updated to mr850alu.Section d4: udi details are not available based on the time of manufacturing.Section g1: contact person updated to mr.(b)(4).Method: the subject mr850 respiratory humidifier (sn (b)(6)), the 900mr805 heater wire adaptor (lot 2101452494) and the 900mr869 temperature/flow probe (lot 140407) in use during the reported event were returned to the fisher & paykel (f&p) healthcare in new zealand where they were visually inspected and functionally tested.Our investigation is based on the information provided by the healthcare facility, f&p healthcare's evaluation of the returned devices, and our knowledge of the products.Results: visual inspection of the mr850 respiratory humidifier identified some minor damage to the enclosure of the mr850 respiratory humidifier.There were no defects or damage observed on the 900mr805 heater wire adaptor or the 900mr869 temperature/flow probe.The mr850 respiratory humidifier and the 900mr869 temperature/flow probe passed functional testing, including the alarm system check.Other than the enclosure damage, no faults or malfunctions were identified with either device.During the functional testing of the 900mr805 heater wire adaptor, an intermittent heater wire alarm occurred during the startup of the subject mr850 respiratory humidifier.Further testing of the 900mr805 heater wire adaptor confirmed an intermittent continuity between one of the heater wire connector pins and its wire.The intermittent connection of the 900mr805 heater wire adaptor would not cause or contribute to the reported elevated temperature on the mr850 respiratory humidifier.A faulty heater wire adaptor would generate an alarm condition on the mr850 respiratory humidifier, and the humidifier will immediately turn off power to the heater wires and heater plate.The healthcare facility did not confirm that any alarm conditions occurred at the time of the reported event.Conclusion: based on the information provided by the healthcare facility and the investigation we have undertaken we have been unable to determine the cause of the reported temperature fluctuation on the mr850 respiratory humidifier or the injury sustained by the patient.The subject mr850 respiratory humidifier and the 900mr869 temperature/flow probe were confirmed to be operating as intended and within specification.The mr850 respiratory humidifier is used to warm and humidify gases delivered to patients requiring mechanical ventilation, positive pressure breathing assistance or medical gases.The mr850 complies with iso8185:2007: "respiratory tract humidifiers for medical use - particular requirements for respiratory humidification systems." this standard specifies safety requirements to prevent the risk of thermal injury, including maximum enthalpy limits for humidified gas delivered to patients through respiratory humidification systems.The mr850 respiratory humidifier features an audible and visual alarm which alerts the user if the displayed temperature exceeds 41°c, or if the airway temperature exceeds 43°c and immediately disables the heater wire and the heater-plate.The mr850 respiratory humidifier user instruction states the following warning: - ensure that both temperature probe sensors are correctly and securely fitted.Failure to do so may result in temperatures in excess of 41°c being delivered to the patient.The rt265 infant evaqua2 breathing circuit user instruction also states the following warnings: do not use heated wire breathing circuits without gas flow.If gas flow is interrupted, turn the humidifier off.Avoid prolonged contact of heated tubes with patient's skin.Check all connections are tight before use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER & PAYKEL HEALTHCARE
Common Device Name
RESPIRATORY HUMIDIFIER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego vargas banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key18323768
Report Number9611451-2023-01126
Device Sequence Number7391422
Product Code BTT
UDI-Device Identifier09420012407245
UDI-Public(01)09420012407245
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberMR850ALU
Device Catalogue NumberMR850ALU
Device Lot Number120403
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2023
Supplement Date Received by Manufacturer03/01/2024
03/01/2024
Initial Report FDA Received Date12/13/2023
Supplement Report FDA Received Date03/28/2024
09/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/03/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
F&P 900MR805 HEATER WIRE ADAPTOR; F&P 900MR805 HEATER WIRE ADAPTOR; F&P 900MR869 TEMPERATURE/FLOW PROBE; F&P 900MR869 TEMPERATURE/FLOW PROBE; F&P RT265 INFANT EVAQUA2 BREATHING CIRCUIT; F&P RT265 INFANT EVAQUA2 BREATHING CIRCUIT; F&P RT265 INFANT EVAQUA2 BREATHING CIRCUIT; GETINGE® SERVO -N VENTILATOR; GETINGE® SERVO -N VENTILATOR; GETINGE® SERVO -N VENTILATOR; HAMILTON MEDICAL® 281976 NCPAP GENERATOR; HAMILTON MEDICAL® 281976 NCPAP GENERATOR; HAMILTON MEDICAL® 281976 NCPAP GENERATOR; HAMILTON MEDICAL® MEDIN NASAL MASK; HAMILTON MEDICAL® MEDIN NASAL MASK; HAMILTON MEDICAL® MEDIN NASAL MASK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age1 DA
Patient SexMale
-
-