| Model Number |
TV-AB2980-N |
| Medical Device Problem Code |
Lack of Effect (4065)
|
| Health Effect - Clinical Code |
Clinical Endoleak (4843)
|
| Date of Event |
11/20/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The patient was implanted with the alto stent graft system to treat an abdominal aortic aneurysm (aaa).A type 1a endoleak was present at the end of this procedure.No intervention was performed and the patient is being monitored.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows that the type 1a endoleak is confirmed.This is consistent with the reported adverse event/incident.Device, user, procedure or anatomy relatedness of this complaint could not be determined.No procedure related harms were identified.The final patient status was reported as discharged home on postoperative day two in stable condition.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Corrections: g3: awareness date has been updated h6: investigation finding codes: remove code 3233 h6: investigation conclusion codes: remove code 11.
|
| |
|
Search Alerts/Recalls
|