SHIRAKAWA OLYMPUS CO., LTD. VISERA HYSTEROVIDEOSCOPE Back to Search Results
Model Number HYF-V
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.In addition, other issues found included: an insufficient curvature angle, discoloration of curved rubber, poor forceps insertion, poor insertion of cleaning brush, lens detached from imaging unit, light guide (lg) bundle broken, uneven lighting, water tightness of control unit and water tightness failure in the curved section, corrosion of actuator, leakage of grip liquid, leakage of up/down (ud) plate liquid, liquid leakage from lg bundle and scratches were found on multiple device components due to handling problem.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event or Problem Description
The customer reported to olympus, the visera hysterovideoscope had buckling of the insertion part.The device was returned for evaluation.During the device evaluation, the metal was found to be protruding from the perforation in the bending section cover due to the breakage of the curved tube.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Additional Manufacturer Narrative
Updated: h6, h10 this report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause of the metal angle wire protruding from the bending section could not be determined, however, the issue was likely the result of observed damage to the bending tube due to user handling/mishandling.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA HYSTEROVIDEOSCOPE
Common Device Name
HYSTEROVIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18325116
Report Number3002808148-2023-14249
Device Sequence Number17981510
Product Code HIH
UDI-Device Identifier04953170340185
UDI-Public04953170340185
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K221557
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHYF-V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2023
Supplement Date Received by Manufacturer12/28/2023
Initial Report FDA Received Date12/14/2023
Supplement Report FDA Received Date01/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/13/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-