| Model Number |
83786 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Stretched (1601); Failure to Advance (2524); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3.Date of event-the event date of (b)(6) 2023 is an estimated based of the aware date of (b)(6) 2023 as the exact date was not reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the coil protruded from catheter tip.The patient was being treated for the mildly tortuous varicose veins in the stomach area.A 12mm x 20cm interlock-35 was inserted through an expect 19 c needle when the coil advanced to the distal part of the needle and got stuck generating a lot of resistance.The entire catheter was removed, but when the coil was pushed out it came out deformed and elongated.The procedure was completed with a different device.No patient complications reported.
|
| |
|