| Model Number |
101/810/080 |
| Medical Device Problem Codes |
Break (1069); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Other text: b3: date of event is unknown; no information has been provided to date.D3, g1,2 email is: regulatory.Responses@icumed.Com.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient had two (2) cannulas delivered, and both were found to be faulty (broken) at opening and unable to be used.No patient injury or clinical affects was reported and the outcome was reported as resolved.(b)(6) and (b)(6) represent the same patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Email is: regulatory.Responses@icumed.Com.H6 - evaluation codes: updated.Device evaluation: one device was returned for investigation.Under visual inspection it was noticed that the inflation line was detached from pilot balloon as reported by customer.The complaint was confirmed.This was attributed to a supplied item fault.This issue will continue to be monitored and further actions taken accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
**udi related data quality updates only**.
|
| |
|