| Model Number |
LSKB1-BX-05 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
11/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Patient was undergoing removal of electrode bolts when two of these fractured during removal, requiring additonal operative maneuvers (incision, partial skull thickness bone removal) to achieve complete device extraction under the same anesthetic.
|
| |
|
Additional Manufacturer Narrative
|
|
Intraoperative x-ray confirmed retained fragments.
|
| |
|
Event or Problem Description
|
|
Patient was undergoing removal of electrode bolts when two of these fractured during removal, requiring additional operative maneuvers (incision, partial skull thickness bone removal) to achieve complete device extraction under the same anesthetic.
|
| |
|
Additional Manufacturer Narrative
|
|
Intraoperative x-ray confirmed retained fragments.---updated 01-02-2024---- the product was returned for evaluation.The alleged complaint was confirmed that the anchor bolt was broken.The anchor bolt broke at the point it was designed to break.Breaking can occur from many things like excessive force when tightening, patient movement (seizures) during monitoring, etc.In this case, the bolts broke at removal but we are not able to determine when the damage leading to the break occurred.The calculated occurrence level matches the occurrence level present in the risk file and the resulting risk level will remain "alap".No further action is needed at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Intraoperative x-ray confirmed retained fragments.Updated 01-02-2024.The product was returned for evaluation.The alleged complaint was confirmed that the anchor bolt was broken.The anchor bolt broke at the point it was designed to break.Breaking can occur from many things like excessive force when tightening, patient movement (seizures) during monitoring, etc.In this case, the bolts broke at removal but we are not able to determine when the damage leading to the break occurred.The calculated occurrence level matches the occurrence level present in the risk file and the resulting risk level will remain "alap".No further action is needed at this time."udi related data quality updates only".Updated updated 09/20/2024.Clarification and correction of the following missing items from previous supplemental report: combination product- no.Brand name- anchor bolt kit, lsb style.Model and catalog number lskb1-bx-05.Primary di number - (b)(4).
|
| |
|
Event or Problem Description
|
|
Patient was undergoing removal of electrode bolts when two of these fractured during removal, requiring additonal operative maneuvers (incision, partial skull thickness bone removal) to achieve complete device extraction under the same anesthetic.
|
| |
|