| Model Number |
SIGTRS60AXT |
| Medical Device Problem Code |
Misfire (2532)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: sigphandle, sig power sigphandle handle (sn:(b)(6)) sigpshell, sig power sigpshell control shell (lot#n3g1445y) sigadaptstnd, sig power sigadaptstnd linear adapter (sn:unknown) h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the sled and staples were visible at the 4cm cut line, and the jaw of the reload was open.Staple pushers were visible at the 5cm cut line.The proximal sutures were cut properly.The distal sutures were returned intact.It was reported that the reload was partially fired and the interlock was engaged.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: failure to completely fire the reload will result in an incomplete cut and/or incomplete staple formation, which may result in poor hemostasis and/or leakage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic pancreatic tail resection, when resecting the pancreas, the jaws were set-up and closed on the pancreas.After that, firing was advanced, but upon advancing by one cm, the firing stopped.Even after pressing any button, nothing started to respond.The screen status of the handle remained stuck at zone two, and there was no change.Both sides of the green button were pressed for 10 seconds, and reboot and a release from the tissue were performed.The tissue has not been stapled, and the release could be performed without any problem.After that, a new powered handle and cartridge were brought in and used, and the surgery was completed.After the procedure, the defective powered stapler and adapter were tested with non-sterile, and it could be used without problem.No patient injury.
|
| |
|
Search Alerts/Recalls
|
|
|