Catalog Number M0031681910
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/03/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Due to the automated mes (manufacturing execution system) system there are controls in the manufacturing process to ensure the product met specifications upon release.During visual inspection, the catheter shaft was found to be kinked/bent at 70cm from the hub tip, the catheter shaft was noted to be blocked/occluded with material 7cm from the hub, and the catheter hub and tip were both intact.During a functional inspection, a 0.023" pin gauge was verified to meet the catheter shaft when inserted into the catheter hub.The microcatheter was flushed, a 0.0158" patency mandrel was advanced through the microcatheter, the mandrel got stuck 7cm from the hub tip, the catheter was cut, and material was removed from the catheter shaft.The material appeared to be ptfe (polytetrafluoroethylene) inner lining from the catheter.The reported event is covered in the device direction for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported defect was confirmed based on analysis of the device.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.Additional information provided by the customer indicated that the device was prepared for use as per the directions for use, there was no damage noted to the packaging prior to opening the packaging, the device was confirmed to be in good condition during preparation/prior to use on the patient and the catheter was flushed to facilitate the guidewire insertion.The catheter shaft was found to be kinked and the patency mandrel could not advance past 7cm into the catheter shaft.There was material removed from the catheter shaft which appeared to be ptfe lining of the catheter.The ptfe would have been damaged when the resistance was encountered during the procedure and during analysis.The catheter shaft was found to be kinked/bent.The as reported 'catheter difficulty advancing guidewire' as well as the as analyzed 'catheter shaft blocked/occluded', 'catheter shaft friction', 'catheter shaft kinked/bent' and 'catheter ptfe inner lining peeling' will be assigned procedural factors as the issue is associated with a product that meets stryker design and manufacture specifications and was used in according with the dfu but due to procedural factors during use, the product performance was limited.
 
Event or Problem Description
The catheter (subject device) was returned for analysis, and it was discovered that the catheter shaft (subject device) was blocked with ptfe (polytetrafluoroethylene) lining from the catheter.The physician replaced it with a new device and continued the procedure without clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXCELSIOR SL-10 PRE-SHAPED 90 150CM
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key18327425
Report Number3008881809-2023-00583
Device Sequence Number7035275
Product Code DQY
UDI-Device Identifier04546540688231
UDI-Public04546540688231
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K042568
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/21/2024
Device Catalogue NumberM0031681910
Device Lot Number23847045
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/06/2023
Initial Report FDA Received Date12/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SYNCHRO GUIDEWIRE (STRYKER)
Patient SexUnknown