Model Number BEA28-120/I16-40
Medical Device Problem Code Lack of Effect (4065)
Health Effect - Clinical Code Clinical Endoleak (4843)
Date of Event 11/28/2023
Type of Reportable Event Serious Injury
Event or Problem Description
The patient was treated for an abdominal aortic aneurysm (aaa) on (b)(6) 2017 with the implant of an afx2 bifurcated stent graft, an afx vela suprarenal, an afx vela infrarenal and an afx limb stent graft.On (b)(6) 2023, it was reported that an indeterminant endoleak was identified.Additional information, including treatment plans and patient status is pending.
 
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.H3 other text : device remains implanted.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident-related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.Endologix was unable to perform an evaluation of the device as it remains implanted in the patient.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the indeterminate endoleak complaint was found to be a type ia endoleak.This is moderately consistent with the reported adverse event/incident.Device, user, procedure or anatomy relatedness of this complaint could not be determined.Procedure related harms for this complaint could not be determined.The final patient status was reported as being monitored.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key18328185
Report Number3011063223-2023-00049
Device Sequence Number20603361
Product Code MIH
UDI-Device Identifier00818009014443
UDI-Public(01)00818009014443(17)180913
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 11/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/13/2018
Device Model NumberBEA28-120/I16-40
Device Lot Number1739764001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/28/2023
Supplement Date Received by Manufacturer01/30/2024
Initial Report FDA Received Date12/14/2023
Supplement Report FDA Received Date02/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/04/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX LIMB STENT GRAFT (LN 1650207012); AFX VELA INFRARENAL (LN 1729374006); AFX VELA SUPRARENAL (LN 1722021008)
Outcome Attributed to Adverse Event Other;
Patient Age76 YR
Patient SexMale
-
-