|
SYNTHES GMBH LARGE QUICK COUPLING FOR TRS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
05.001.213 |
| Medical Device Problem Codes |
Mechanical Jam (2983); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was traced to component failure due to normal wear.Udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported from netherlands that during service and evaluation, it was determined that the large quick coupling device was worn and would not move.It was further determined that the device failed pretests for general condition, check the tool coupling, check for mechanical free movement and check general function in running mode.It was noted in the service order that the device had an unspecified malfunction.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|