| Model Number |
100145CO2EXT |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An in-depth investigation was conducted and determined that the cause of the tubing not staying clamped was attributed to the tip (point) of the clip.The tip is seated within the "teeth" of the clip to keep the clip engaged/closed.As the tip was not sharp/pointy enough, this allowed the tip to detach from the teeth resulting in the reported issue.The supplier was made aware of the complaints associated with this issue.Medivators will continue to monitor for similar events to ensure the product continues to perform as expected.
|
| |
|
Event or Problem Description
|
|
The user facility reported that their endo smartcap tubing pinch clamp will not stay clamped and causes water to leak out.No report of injury or procedure delay.
|
| |
|
Search Alerts/Recalls
|