EXACTECH, INC. NV GXL LNR, NEUTRAL, 32MM ID, GROUP 3 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Model Number |
NV GXL LNR, NEUTRAL, 32MM ID, GROUP 3 CUPS |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Osteolysis (2377)
|
| Date of Event |
11/08/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported by the germany competent authorities, the patient presented for several follow-up checks with persistent symptoms on the opposite side.The x-ray control revealed a clear decentering of the prosthetic head.The ct scan also showed osteolysis in the acetabulum as a sign of premature inlay wear.This was verified when the inlay was changed on (b)(6) 2023 to a vitd-hardened, specially approved inlay (novation xle, neutral liner group 3, 36 mm i.D).As part of the replacement operation, in addition to the inlay replacement, the firm socket integrity was determined as well as the curettage and sealing of the cysts in the socket using allogeneic cancellous bone and the replacement of the prosthetic head.The explants are currently in the laboratory doctor's office for microbiological examination using sonication.After the examination (14 days of long-term incubation).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
The prosthesis wear reported may have been due to a combination of the risk factors specified in the hhe, such as use error, implant positioning, implant size selection, patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential), the use of nylon vacuum bags without evoh, surgical approach, or any combination of these possibilities.However, this cannot be confirmed since the devices were not returned for evaluation and radiographs were not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: h6, h11.The reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.A definitive root cause was unable to be determined; however, may have resulted from use error, implant positioning, implant size selection, patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential), the use of nylon vacuum bags without evoh, surgical approach, or any combination of these possibilities.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|