Model Number FX802T
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 11/20/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that an m.Blue 10 valve (#fx802t) was implanted during a procedure performed on an unknown date.According to the complainant, the shunt showed a fault after implantation.The patient underwent a revision procedure performed on (b)(6) 2023.No patient complications were reported as a result of the revision procedure.The adverse event is filed under aic reference xc (b)(4).
 
Additional Manufacturer Narrative
Investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
Additional information: corrected information: b3(date of event), b5 (revision date updated).Visual inspection: during the investigation, no significant deformations or damage of the valve was determined.Permeability test: a permeability test has shown that the m.Blue is permeable.During the permeability test bloody liquid were flushed out.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.The results show that the m.Blue operates within the accepted tolerance in either position.Adjustment test: the m.Blue valve was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection: after dismantling of the valve, no deposits were found in the m.Blue to make the deposits in the valve more visible, they were colored using a staining solution.Results: based on our investigation results, we cannot determine functional deviations or other abnormalities in the product.How the abovementioned functional impairment occurred is not clear to us at the time of examination.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a m.Blue plus (#fx802t) was implanted during a procedure performed on (b)(6) 2023.According to the complainant, the valve showed an under-drainge.The patient underwent a revision procedure performed on (b)(6) 2023.The complainant device has returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 3 years, 3 months.Weight: 16 kgs.Height: 97 cm.Gender: male.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.BLUE
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
MDR Report Key18330636
Report Number2916714-2023-00129
Device Sequence Number18255579
Product Code JXG
UDI-Device Identifier04041906504410
UDI-Public4041906504410
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/27/2023
Report Date (Section F) 12/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX802T
Device Catalogue NumberFX802T
Device Lot Number20062188
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date11/20/2023
Device Age13 MO
Event Location Hospital
Date Report to Manufacturer11/20/2023
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/14/2023
Supplement Report FDA Received Date12/27/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age3 YR
Patient SexMale
Patient Weight16 KG
Date Report Sent to FDA12/14/2023
-
-