BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-35 10/80/130; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
|
Back to Search Results |
|
| Model Number |
383271 |
| Medical Device Problem Code |
Failure to Unfold or Unwrap (1669)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the affected device nor the angiographic material was returned.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each device is inflated and tested for air tightness by means of a pressure test and a helium leakage test.Based on the conducted investigations, no manufacturing or material related root cause could be determined.
|
| |
|
Event or Problem Description
|
|
A passeo-35 balloon catheter was selected for treatment.The balloon catheter was positioned in the target lesion, but could only be inflated up to 7-9 atm, not fully inflated.
|
| |
|