Model Number 2120
Medical Device Problem Codes Device Alarm System (1012); Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/16/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Other text: a supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.D3, g1, and g2 email is: (b)(6).
 
Event or Problem Description
It was reported that the pump exhibited an air in line alarm.Troubleshooting was performed and the pump continued to alarm.No patient injury was reported.
 
Additional Manufacturer Narrative
No product was returned.We are unable to confirm the reported complaint.If the product is returned, the manufacturer will reopen this complaint for further investigation.The most probable cause is air detector sensor; however, this cannot be confirmed as no product was returned for investigation.The service history review identified no indication that the complaint was related to a service of the device within the review period.D3, g1,g2 email is: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-SOLIS AMBULATORY INFUSION PUMP
Common Device Name
PUMP, INFUSION
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
reed covert
MDR Report Key18331181
Report Number3012307300-2023-11067
Device Sequence Number18246593
Product Code FRN
UDI-Device Identifier15019517096392
UDI-Public15019517096392
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K111275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2120
Device Catalogue Number21-2120-0103-78
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2023
Supplement Date Received by Manufacturer12/28/2023
Initial Report FDA Received Date12/14/2023
Supplement Report FDA Received Date01/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/14/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-