• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #2; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #2; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 6021-0230
Medical Device Problem Codes Material Erosion (1214); Unstable (1667)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994); Metal Related Pathology (4530)
Date of Event 09/29/2008
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.H3 other text : device not returned.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head and accolade tmzf plus hip stem on her right hip on or about (b)(6) 2008.It is further alleged that after implantation she experienced pain and suffering and loss mobility and required a revision surgery.
 
Additional Manufacturer Narrative
Reported event: an event regarding altr and wear involving an accolade stem that was mated with an lfit v40 cocr head was reported.The event of altr was confirmed.Method & results: -product evaluation and results: visual inspection, material analysis, functional, and dimensional inspections could not be performed as the device was not returned.-clinician review: a review with a clinical consultant indicated conclusion/assessment: the patient underwent a right tha in 2008 for oa.She underwent a left tha for djd but potential infection.The first hip had a tmzf stem the second an accolade 2, both with a v40 head.The right hip was revised for reported increased metal levels and pseudotumor on 09/10/2015.At the time of surgery metallosis, pseudotumor and abductor disruption were noted.Medical records before or after the surgery, radiographs, imaging studies, and/or laboratory studies were not provided for review.Event confirmation: a right tha can be confirmed.A left tha can be confirmed.A revision right tha can be confirmed.Pseudotumor and abductor involvement can be confirmed at the revision right tha.Metallosis, increased metal levels, pain, suffering, and/or loss of mobility cannot be confirmed without additional information and records.Root cause: the root cause of the right tha was oa.The root cause of the left hip djd and tha is unclear.The root cause of the revision right hip was clunking and pain and reported increased metal levels (not documented).The other findings could not be confirmed.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to altr, wear (metallosis/trunnionosis).The reported accolade stem was mated with an lfit v40 cocr head.The cocr head has been identified to be within scope of an nc and capa.No further investigation for this event is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head and accolade tmzf plus hip stem on her right hip on or about (b)(6) 2008.It is further alleged that after implantation she experienced pain and suffering and loss mobility and required a revision surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE PLUS TMZF HIP STEM #2
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key18331993
Report Number0002249697-2023-01564
Device Sequence Number6938067
Product Code LPH
UDI-Device Identifier04546540510679
UDI-Public04546540510679
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2015
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2013
Device Catalogue Number6021-0230
Device Lot Number26919904
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/27/2023
Supplement Date Received by Manufacturer04/29/2024
Initial Report FDA Received Date12/14/2023
Supplement Report FDA Received Date04/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/29/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age59 YR
Patient SexFemale
-
-