The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd module with drive pcb.In addition, our technician confirmed that the angle wire fluid damage, the insertion flexible tube fluid damage, the segment fluid damage, the control body fluid damage, the up pulley wire fluid damage, the left pulley wire fluid damage, the lcb (light carrying bundle) fluid damage, the us connector cable (long) buckled, the us connector/junction cable (short) buckled, the operation channel (primary) leak, the operation channel (primary) cut, the warning/caution label worn out, the angulation down angulation decrease, and the angulation left angulation decrease; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|