|
Procedure/medical information review: imaging review, (b)(6) , md, (b)(6) 2023: eleven radiographic images are supplied, and embedded in a word file.They are not labeled as to date or time.Image quality is poor.Image 1: 3d dsa (transparent vessel mode) of right ica.There is a medium-sized pcom aneurysm.Some measurements are included on the image, but they are blurry and difficult to read.Width at dome: ?6mm?, difficult to read.Width at base: 8.02 mm.Depth at three different points: 5.04, 4.29, and 4.73 mm.Image 2: ap right ica dsa, subtracted, contrast.The same pcom aneurysm is demonstrated.Image 3: lateral right ica dsa, subtracted, contrast.The same pcom aneurysm is demonstrated.Image 4: ap oblique right ica dsa, subtracted, contrast.The same pcom aneurysm is demonstrated.A guiding catheter is in the distal petrous ica.A microcatheter and microguidewire are in the pre-cavernous ica, below the aneurysm.Image 5: ap oblique (slightly different than image 4) right ica dsa, subtracted, contrast.The same pcom aneurysm is demonstrated.A guiding catheter is in the distal petrous ica.A microcatheter and microguidewire are in the pre-cavernous ica, below the aneurysm.Image 6: ap oblique (slightly different than image 5) right ica dsa, subtracted, contrast.The same pcom aneurysm is demonstrated.A guiding catheter is in the distal petrous ica.A microcatheter and microguidewire are in the pre-cavernous ica, below the aneurysm.Image 7: ap oblique right ica dsa, subtracted, contrast.A via catheter has been placed in the aneurysm.Image 8: ap oblique right ica dsa, unsubtracted, contrast.A via catheter is still in the aneurysm.Its tip may be slightly outside the aneurysm, but it is difficult to tell on this image.Image 9: ap oblique (slightly different than image 8) right ica dsa, unsubtracted, contrast.A via catheter is still in the aneurysm.Its tip may be slightly outside the aneurysm, but it is difficult to tell on this image.Image 10: ap oblique right ica radiograph, unsubtracted, no contrast.The via catheter is still in the aneurysm.Another microcatheter is now in the aneurysm, being used to deliver coils (a moderate coil pack is in place inside the aneurysm).Image 11: ap oblique right ica radiograph, unsubtracted, no contrast.No change, except there is a microcatheter with two distal markers (likely a balloon microcatheter) positioned around the expected position of the aneurysm neck (balloon not inflated).The microguidewire is in the distal m1.The images do not provide an explanation relating to the complaint regarding the reported perforation of the aneurysm.Investigation findings: items returned: - n/a.Visual analysis: a visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation.Investigation findings: without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.Based on a review of the device¿s risk documentation, the reported event did not indicate there were any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform and analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Ifu review (additional information can be found in the ifu): potential complications.Potential complications include but are not limited to the following: hematoma at the site of entry, aneurysm rupture, emboli, vessel perforation, parent artery occlusion, hemorrhage, ischemia, vasospasm, vascular thrombosis and neurological deficits including stroke and death.Warnings never advance or withdraw a device against resistance until the cause of the resistance is determined by fluoroscopy.Excessive force against resistance may result in damage to the device or vessel perforation.When injecting contrast for angiography, ensure the catheter is not kinked or occluded.Do not exceed 300 psi maximum recommended infusion pressure.Excess pressure may result in catheter damage or patient injury.Steam shaping may result in improper device delivery and deployment, depending on the degree of shaping and catheter deflection during device delivery.Using the via microcatheter with guide catheters smaller than what is recommended (see compatability table above) may result in damaging the hydrophilic coating.Precautions the via microcatheter has a lubricious hydrophilic coating on the outside of the catheter.It must be kept hydrated in order to be lubricious.This can be accomplished by attaching a y-connector to a continuous saline drip.The operator should be aware that microcatheters, in distal blood vessels, may increase the risk of thromboemboli.If removed from the patient, the hydrophilic coating on the via microcatheter should be hydrated with heparinized saline.Do not allow the coating to dry as this may impact the coating safety and performance.Avoid wiping the device with dry gauze as this may damage the device coating.Avoid excessive wiping of the coated device.Procedure.Catheterization of the lesion.5.If desired, carefully introduce the microcatheter through the introducer sheath.6.Prepare an appropriately sized guidewire and insert into the microcatheter per the manufacturer¿s instructions.7.Introduce the guidewire and microcatheter as a unit into the guiding catheter until the distal tip of the guide catheter has been reached.Alternatively advance the guidewire and microcatheter until the desired site has been reached.Verify catheter position using fluoroscopy.Gently tighten rhv, as needed, without crushing the microcatheter.8.Peel away introducer sheath by pulling on the tab, if used.9.After the microcatheter has been positioned inside the lesion, remove the guidewire.10.Flush the id of the microcatheter with sterile flush solution by attaching a syringe to the catheter hub.11.Attach a second rhv to the hub of the microcatheter.Attach a one-way stopcock to the sidearm of the second rhv and connect the flush solution line to the stopcock.12.Open the stopcock to allow flush through the microcatheter with sterile flush solution.Removal of the via microcatheter.13.Under fluoroscopic guidance, withdraw the via microcatheter until the entire device has been removed from the patient.
|
|
Upon review of device registrations and mdrs submitted to the fda, discrepancies were noted.Clarification was received regarding the device registration and the following fields have been corrected: d1, d2a, d2b- correct, no change, d3, d4, di number correct, no change, g4, g5.The corrected information does not have any impact on the incident or investigative data submitted on earlier reports.The h6 codes and h11 investigation results stand as previously submitted.
|