• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. MISSION KIT; BIOPSY INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. MISSION KIT; BIOPSY INSTRUMENT Back to Search Results
Catalog Number 1820MSK
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/23/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that prior to a biopsy procedure, the coaxial needle in the kit allegedly was not compatible with the biopsy needle.It was further reported that the internal diameter of the coaxial cannula allegedly may be smaller or larger than the external diameter of the biopsy needle.There was no patient contact.
 
Additional Manufacturer Narrative
H10: a voluntary recall has been initiated for mission disposable core biopsy instrument kit which are product catalog/lot number specific.Reportedly the mission disposable core biopsy instrument kit may have a mismatch between the coaxial and the needle in the kit devices.Device incompatibility due to the external diameter of the biopsy instrument being larger than the internal diameter of the coaxial needle is most likely to result in no patient involvement or a clinically insignificant procedural delay.If the coaxial needle has been deployed, an additional device would be required.If a compatible device is not immediately available, the procedure may need to be repeated, resulting in minor injury from a new needle puncture and care may be prolonged.If efforts are made to use the device despite incompatibility, unusual adverse events such as needle tip breakage may occur.To date there have been no adverse events worldwide reported related to this issue.As result of the field action, this event is being reported as a malfunction reportable event.H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiration date: 03/2025) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MISSION KIT
Common Device Name
BIOPSY INSTRUMENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe 85281
Manufacturer (Section G)
CAREFUSION D.R. 203 LTD.
zona franca las americas
km 22-e-1
santo domingo 11606
DR   11606
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18333632
Report Number2020394-2023-01263
Device Sequence Number12874537
Product Code KNW
UDI-Device Identifier00801741097072
UDI-Public(01)00801741097072
Combination Product (Y/N)N
PMA/510(K) Number
K171953
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 12/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1820MSK
Device Lot Number0001461273
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/23/2023
Initial Report FDA Received Date12/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
UNKNOWN
Patient Sequence Number1
Patient SexUnknown
-
-