• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; DATA MANAGEMENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; DATA MANAGEMENT SYSTEM Back to Search Results
Model Number 71926-01
Medical Device Problem Code Device Alarm System (1012)
Health Effect - Clinical Codes Hypoglycemia (1912); Loss of consciousness (2418)
Date of Event 11/27/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Extended investigation is pending at this time.A follow up will be submitted once additional information is obtained.The device mfg date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
An alarm issue was reported with the abbott diabetes care (adc) device in use with iphone se phone with ios operating system version 16.5.1.The low and high glucose alarms did not sound and customer was not alerted of changes in glucose level.As a result, the customer experienced a loss of consciousness, and was unable to self-treat, requiring third-party treatment of "re-sugared" by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.
 
Additional Manufacturer Narrative
The freestyle librelink app complaint was investigated and determined that there were no issues with the librelink application that would have led to the complaint.The user reported missing high or low glucose alarms.Attempted to replicate the user¿s complaint using similar configuration and the reported issue was unable to be replicated and the system functioned as intended.Therefore, this complaint is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
An alarm issue was reported with the abbott diabetes care (adc) device in use with iphone se phone with ios operating system version 16.5.1, app version 2.10.2.7677.The low and high glucose alarms did not sound and customer was not alerted of changes in glucose level.As a result, the customer experienced a loss of consciousness, and was unable to self-treat, requiring third-party treatment of "re-sugared" by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.
 
Event or Problem Description
An alarm issue was reported with the abbott diabetes care (adc) device in use with iphone se phone with ios operating system version 16.5.1, app version 2.10.2.7677.The low and high glucose alarms did not sound and customer was not alerted of changes in glucose level.As a result, the customer experienced a loss of consciousness, and was unable to self-treat, requiring third-party treatment of "re-sugared" by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.
 
Additional Manufacturer Narrative
The freestyle libre 2 app complaint was investigated and determined that there were no issues with the libre 2 application that would have led to the complaint.The user reported missing high or low glucose alarm.Attempted to replicate the user¿s complaint using similar configuration of iphone 11 pro, ios 16.6, 2.10.2.7677 and the reported issue was unable to be replicated and the system functioned as intended.Therefore, this complaint is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Common Device Name
DATA MANAGEMENT SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key18333868
Report Number2954323-2023-54083
Device Sequence Number4479235
Product Code PZE
UDI-Device Identifier00357599810009
UDI-Public00357599810009
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P160030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/19/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number71926-01
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2023
Supplement Date Received by Manufacturer01/07/2024
03/11/2024
Initial Report FDA Received Date12/15/2023
Supplement Report FDA Received Date01/15/2024
03/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/07/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-