| Model Number |
71926-01 |
| Medical Device Problem Code |
Device Alarm System (1012)
|
| Health Effect - Clinical Codes |
Hypoglycemia (1912); Loss of consciousness (2418)
|
| Date of Event |
11/27/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Extended investigation is pending at this time.A follow up will be submitted once additional information is obtained.The device mfg date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
An alarm issue was reported with the abbott diabetes care (adc) device in use with iphone se phone with ios operating system version 16.5.1.The low and high glucose alarms did not sound and customer was not alerted of changes in glucose level.As a result, the customer experienced a loss of consciousness, and was unable to self-treat, requiring third-party treatment of "re-sugared" by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The freestyle librelink app complaint was investigated and determined that there were no issues with the librelink application that would have led to the complaint.The user reported missing high or low glucose alarms.Attempted to replicate the user¿s complaint using similar configuration and the reported issue was unable to be replicated and the system functioned as intended.Therefore, this complaint is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
An alarm issue was reported with the abbott diabetes care (adc) device in use with iphone se phone with ios operating system version 16.5.1, app version 2.10.2.7677.The low and high glucose alarms did not sound and customer was not alerted of changes in glucose level.As a result, the customer experienced a loss of consciousness, and was unable to self-treat, requiring third-party treatment of "re-sugared" by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.
|
| |
|
Event or Problem Description
|
|
An alarm issue was reported with the abbott diabetes care (adc) device in use with iphone se phone with ios operating system version 16.5.1, app version 2.10.2.7677.The low and high glucose alarms did not sound and customer was not alerted of changes in glucose level.As a result, the customer experienced a loss of consciousness, and was unable to self-treat, requiring third-party treatment of "re-sugared" by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The freestyle libre 2 app complaint was investigated and determined that there were no issues with the libre 2 application that would have led to the complaint.The user reported missing high or low glucose alarm.Attempted to replicate the user¿s complaint using similar configuration of iphone 11 pro, ios 16.6, 2.10.2.7677 and the reported issue was unable to be replicated and the system functioned as intended.Therefore, this complaint is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|