Pma/510(k) # p100022/s014.Device evaluation: the zisv6-35-125-7-60-ptx device of lot number c2098020 involved in this complaint was implanted in the patient and was not available for evaluation.With the information provided, a document-based investigation was conducted.Manufacturing records: prior to distribution all zilver devices are subjected to a visual inspection and functional inspection to ensure device integrity.A review of the manufacturing records did not reveal any discrepancies that could have contributed to this complaint issue.Instructions for use/label: it should be noted that the instructions for use (ifu0118) states the following: ¿the zilver ptx drug eluting peripheral stent is indicated for improving luminal diameter for the treatment of de novo or restenotic symptomatic lesions in native vascular disease of the above the knee femoropopliteal arteries".There is evidence to suggest that the customer did not follow the instructions for use, the device was used in the right common iliac.Image review: an image was not returned for evaluation.Root cause analysis: a definitive root cause of off label use was determined from the available information.It is known that the zilver ptx stent was used in the right common iliac.This is not in line with the intended use of the device.As previously mentioned, the ifu states the product is intended for use in the treatment of de novo or restenotic symptomatic lesions in native vascular disease of the above the knee femoropopliteal arteries.This off label use of the device, i.E., the diameter of the stent would be smaller than that of the common iliac, which likely caused the device landed further up into the aorta and the stent wouldn¿t be imposed well.Off label use complaints are considered to be unforeseen misuse.It is unknown how the device will function outside of its intended use.Confirmation of complaint: complaint is confirmed based on visual and/or functional inspection.Summary: according to the initial reporter, the intended target site for the device was the right common iliac; however, the device landed further up into the aorta as the user believed it was not well imposed.The user wanted to use a cuff to rectify but the stent was very mobile so they elected to leave it alone.It appeared it would migrate north toward the renal so they elected to not intervene.Confirmed quantity of 1 device, confirmed prior to use.According to the initial reporter, a section of the device did remain inside the patient¿s body.An attempt was made on the same day to capture vs exclude but the user was unable to retrieve/ exclude the stent.It appeared the stent would migrate north toward the renal so they elected to not intervene.Investigation findings conclude a definitive root cause of off label use was determined.The user has not complied with the requirements of the instructions for use with respect to the intended use of the device.Off label use complaints are considered to be unforeseen misuse.It is unknown how the device will function outside of its intended use.As the device was used outside of their validated state and/or against the instructions provided in the ifu, it is not possible to predict how the devices will perform or function.Complaints of this nature will continue to be monitored for potential emerging trends.
|