ROCHE DIAGNOSTICS ELECSYS SYPHILIS; ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM Back to Search Results
Catalog Number 07251378190
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/24/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The cobas e 801 analytical unit serial number was (b)(6).The investigation is ongoing.
 
Event or Problem Description
There was an allegation of a questionable syphilis elecsys e2g result from the cobas e 801 analytical unit.The result from the cobas e801 was 0.0892 (negative).The result by trust method (shanghai rongsheng) was weak positive 1:1.At another site, the rpr result was weak positive 1:1.
 
Additional Manufacturer Narrative
The investigation determined the issue was consistent with a sample-specific discrepancy.The elecsys syphilis assay performed within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS SYPHILIS
Common Device Name
ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18335098
Report Number1823260-2023-04027
Device Sequence Number8407005
Product Code LIP
UDI-Device Identifier04015630940356
UDI-Public04015630940356
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K160910
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Catalogue Number07251378190
Device Lot Number682565
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/28/2023
Supplement Date Received by Manufacturer12/22/2023
Initial Report FDA Received Date12/15/2023
Supplement Report FDA Received Date12/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-