Model Number 20E
Medical Device Problem Code Device Sensing Problem (2917)
Health Effect - Clinical Code Asystole (4442)
Date of Event 11/21/2023
Type of Reportable Event Malfunction
Event or Problem Description
The customer contacted stryker to report that during a patient event their device was unable to detect the patient through the defibrillation electrodes.The customer advised that the device prompted them to ¿connect electrodes¿ when the defibrillation electrodes were connected to the patient.The customer advised they were using third party conmed padpro defibrillation electrodes.The patient did not survive.The customer reported that it was considered likely that the patient was found in asystole, or expired, but the paramedics were performing required measures upon their delivery of the patient to the facility.
 
Additional Manufacturer Narrative
The customer provided stryker with all the available patient information.Patient fields in which information was not provided were intentionally left blank.Stryker performed a clinical review of the reported issue and determined that the device use did not contribute to the patient outcome due to the following: the healthcare provider reported that the patient was in asystole or expired before the resuscitation began.In the medical judgment of these reviewers, the device did not cause or contribute to the reported outcome.Stryker continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Additional Manufacturer Narrative
Stryker evaluated the customer's device and was able to verify the reported issue through the electronic records of device.However, the reported issue could not be duplicated.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determined.
 
Event or Problem Description
The customer contacted stryker to report that during a patient event their device was unable to detect the patient through the defibrillation electrodes.The customer advised that the device prompted them to ¿connect electrodes¿ when the defibrillation electrodes were connected to the patient.The customer advised they were using third party conmed padpro defibrillation electrodes.The patient did not survive.The customer reported that it was considered likely that the patient was found in asystole, or expired, but the paramedics were performing required measures upon their delivery of the patient to the facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 20E DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
brian blakeslee
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key18335199
Report Number0003015876-2023-02376
Device Sequence Number20518187
Product Code MKJ
UDI-Device Identifier00883873852804
UDI-Public00883873852804
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P160026
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number20E
Device Catalogue Number99507-000132
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/21/2023
Supplement Date Received by Manufacturer03/19/2024
Initial Report FDA Received Date12/15/2023
Supplement Report FDA Received Date03/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/28/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age51 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-